ZDJ-15KArc Resistance Tester

Suitable for batch production inspection of insulating tools and equipment such as gloves, boots, rods, and voltage testers, it can reliably identify parameters including leakage current, insulation aging, and power frequency withstand voltage. It can also perform voltage withstand tests on insulating blankets, mats, and other similar tools.


  • 产品型号:ZDJ-15KArc Resistance Tester
  • 产       地:Changchun City, Jilin Province
  • 更新时间:2025-04-02
  • 访  问  量:24
产品图册

products

Compliant standards

GB/T1411-2002Dry,solid insulating materials-Resistance test to high-voltage, low-current arc discharges》 (Equivalent to IEC 61621-1997and ASTM D495 and other standardscomply UL 746Aand other test methods.

Application

NDH-15K Arc resistance test system is a high precision equipment specially developed for the test of electrical insulation materialsits suitable for electric insulation materials, plastics, film, resin, mica, ceramics, glass, insulating paint and other insulating materials arc resistance evaluation,it can also be used in product testing and research of electrical properties of new materials.

Functional principle

Insulation materials have the ability to resist deterioration caused by high voltage arc actionusually a standard arc flame is used to induce carbonization on the surface of the material until the surface is conductive.The time required for the arc to disappear is expressed in second. Add two tungsten electrodes separated by a certain distance on the plate sample, at first the arc is intermittent, then gradually shortened,and finally continuous arc with gradually increasing current,the material is subjected to increasingly severe arc conditions and eventually fails.Arc resistance is important for insulating materials used in high-voltage switches or low-voltage high-current switches,because the switch is often affected by arc when it is opened and closed, only materials with good arc resistance can be selected.

Insulating material used in the interior of electrical and electronic equipment, when exposed to a high-voltage, low-current arc discharge, is tested the differences in surface damage close to the arc.The arc discharge causes local thermal and chemical decomposition and corrosion and eventually forms a conductive channel on the insulating materialWhen a conductive channel is formed in the tested material, the material can be determined to have failed.

    Arc resistance tester is GB1411, IEC61621, ASTMC495, UL746A and other standards of the simulation safety test project necessary special instrument.

The arc resistance tester is composed of electrical schematics and electrodes according to the following standard requirements

Structure and Composition

The system consists of two parts

1Control section:The lower part of the arc resistance tester is the operating systeminternal electrical control system, test can be performed automatically by selecting a program,and the digital display of each test process and results, test start, stop, breakdown and trip protection, test termination judgment.

2High pressure test section:It consists of electrode system

微信图片_20250427141918.png

Figure 1

The electrode material is tungsten roddiameter 2.4mm±0.05mm,

The Angle between the ellipse tip surface and the axis is 30°±1°

A force of 0.5N±0.05N is applied to the sample

The two electrodes are tilted horizontally 35°±1°The Angle between the axes is 110°±2°),

Two tip spacing6.35±0.1mmIEC                  6.0±0.1mmASTM

This product uses single chip microcomputer and touch screen control, high degree of automation, can automatically regulate the voltage,automatically switch the execution program, automatically describe the running curve, measure the real-time voltage and current, and time the test time.

The electrode system imitated the German imported electrode, Angle, gap calibration is convenient, fast and efficient, it has good repeatability and high precision

Data measurement using high precision, high-speed conversion chip, fast conversion speed, high measurement precision, strong anti-interference ability.

微信图片_20250427141933.png 

Figure 2

With the shield closed, the voltage is applied and the current is regulated by variable resistors R10, R20, R30 and R40 to observe whether the sample forms a conductive channel. See figure1 R10R20R30R40.

Scope of application

PCB

Photovoltaic film materials

Ptfe, ABS, nylon and other plastics

Film material

Insulating paint

Mica materials

Ceramic, glass fiber, etc

Equipment parameter

1Arc on/off time error: All program segments are superior±1ms

2Voltage regulator capacity1kVA

3Rated test voltage12.5kV

   Maximum test voltage15kV

4Test voltage accuracy:<1.5%

5Current measurement accuracy:<1.5%

6Electrode pressure on sample0.5±0.05N

7Electrode materialTungsten bars

8Distance between electrodes6.35±0.1mmIEC 6.0±0.1mmASTM

9Ambient temperature:(23±2

10Ambient humidity:(50±5%

11Power supply220V±5%  50Hz  10A

12 Equipment weight58kg                          Packed weight80kg

13 Equipment dimension:(L*W*H)670*660*590mm  Packed dimension780*770*830mm

 

Program segment

Arc current

mA

Arc onoff time /s

Duration of each program segment/s

Total test time/s

IEC 61621

ASTM D495

1

10

1/8son 7/8s off

1/8son 7/8s off

60

60

2

10

1/4s on 3/4s off

1/4s on 3/4s off

60

120

3

10

1/4s on 1/4s off

1/4s on 1/4s off

60

180

4

10

Continuous power on

60

240

5

20

Continuous power on

60

300

6

30

Continuous power on

60

360

7

40

Continuous power on

60

420

Standard Configuration

1  Arc resistance tester host machine 1

2  Electrode system  1 set

3  Power cable 1

4  Standard electrode spacing measure block 1

5  Angle standard block 1

6  Product Ex-factory report 1

7  Sample stand 1

8  Operation instruction 1

9  Packing list 1

NoteAll instruments and equipment need to be grounded and the grounding resistance is ≤4Ω. The supply and demand sides agree that the supplier should provide technical requirements and the demander should build the grounding system by itself.